Cell counting techniques

Diluted Specimens

Calibrated automatic pipets should be used on specimens that require a dilution. Mouth pipetting must of course always be avoided.

Automated cell counters may not be the best choice for cell counts on CSF because of the variation in background counts.

A high background count could cause a false increase in normal or slightly elevated count. The dilution required is based on the appearance of the sample.

AppearanceDilution RatioVolume of SampleVolume of Diluent
slightly hazy1:1030 ul270 ul
hazy1:2030 ul570 ul
slightly cloudy1:10030 ul2970 ul
slightly bloody1:10030 ul2970 ul
cloudy1:20030 ul5970 ul
bloody1:20030 ul5970 ul
turbid1:100000.1 ml of a 1:100 dilution9.9 ml